SoloVPE Header

Frequently Asked Questions

Q: What is Slope Spectroscopy?
A: Slope Spectroscopy TM is at its heart of UV spectroscopy method based upon the Beer-Lambert Law. However, if differs from traditional spectroscopy methods which hold the measurement pathlength constant because it varys the measurement pathlength to collect Absorbance vs. Pathlength or Section data. The Slope of the Section plot allow rapid quantitation of concentrations and because the measurement pathlength can change based upon the sample concentration and absorbance, highly accuract measurements can be made without the need or cost of dilution steps. To learn more read this article or Wiki.


Q: What concentration ranges can the SoloVPE measure?
A: The SoloVPE System is capable of measureing a wide range of concentrations, including Protein Concentrations (e.g. BSA) from below 0.01 mg/ml to over 300 mg/ml and Antibody Concentrations between 0.01 mg/ml to 250 mg/ml.


Q: How much sample volume is required to make a measurement?
A: The amount of sample required for a measurement is dependent on the Sample Vessel being used and the concentration (which influences the pathlengths to be measured). Highly concentrated samples can be measured with less than 10 microliters, however, very dilute samples may required more than 1 ml. In each case the sample can be recovered


Q: How long does a concentration measurement take?
A: The fastest and easiest way to measure concentration with the SoloVPE System is to use the Quick Slope application. Because Baseline Correction is frequently unnecessary since the Slope Spectroscopy technique is used, measurement results can typically be obtain in about 1 minute.


Q: Can the SoloVPE be used with any spectrophotometer?
A: The SoloVPE System can only be used with the Varian Cary 50 or Agilent Cary 60. The unique design and capabilities of those instruments make them ideal for integrating variable pathlength technology.


Q: Can the SoloVPE System be used in a GMP Environment?
A: SoloVPE Systems have been installed and validated in GMP labortatory environments around the globe. When being installed in a Quality Assurance lab, customers typically purchase the option security package upgrade call SecureVPE.


Q: Where are SoloVPE System currently being used?
A: Many companies around the world have purchased, installed and validated SolovPE Systems. System are installed in the Americas, Europe and Asia. Many companies have standardized on the SoloVPE System across multiple sites for concentrations measurements because of the excellent accuracy and reproducibility of the results. SoloVPE System have been installed across every functional area from Discovery and R&D through pilot production, manufacturing and quality assurance.